Updated:2023-09-06
Views:1046
Online Service
The ViSolid is a reflectance-based TSS (total suspended solids) probe for the IQ SensorNet process monitoring system.
The ViSolid comes with 2 factory calibrations for greatest versatility. Calibration matrix type 1 is designed for monitoring MLSS (mixed liquor suspended solids), return activated sludge (RAS), and waste activated sludge (WAS). Calibration matrix type 2 is used for monitoring primary sludge and thickened sludge. Two types of user calibrations (correction factor and value pair) are available for fine-tuning the factory calibration or monitoring TSS in other applications.
Smooth, durable sapphire measuring windows and continuous ultrasonic cleaning provide reliable and trouble-free service. The ViSolid is a good choice for process control of activated sludge wasting with an immersion mount for MLSS monitoring and an insertion mount to monitor RAS/WAS.
Factory calibrated
UltraClean™ maintenance free cleaning system provides integrated, continuous ultrasonic cleaning to prevent fouling for accurate measurements every time.
Low ownership cost. No wipers or seals to replace.
Measure in mg/L or %. Measuring range up to 40,000 mg/L
2-year probe warranty
Anti Fouling | UltraClean™ ultrasound cleaning system |
Certifications | UL 61010-1, CAN/CSA C22.2 # 61010.1 |
Depth / Pressure Rating / Limit | 10 bar (145 psi) |
Equipment used with | 282/284, 2020 3G |
Integrated Lightning Protection | Yes |
Languages | English, German |
Measurement Method / Channels | in case of higher concentrations, a scattered light method is used, the backscatter method is implemented |
Measurement Range | Concentration: 0 to 400 mg/L; 0 to 4000 mg/L; 0 to 25,000 mg/L; 0 to 40,000 mg/L Saturation: 0.0003 to 100% |
Operating Temperature | 32 – 140 °F (0 – 60 °C) |
Parameters Measured | Total Suspended Solids |
Precision | Repeatability according to DIN ISO 5725 or DIN 1319 <0.015% or > 0.0006 FNU |
Sampling | Yes |
Smart Sensors / Ports | Yes |
Storage Temperature | 23 – 149 °F (-5 – +65 °C) |
Unit of Measure | g/l SiO2 / % SiO2, g/l TSS / % TSS |
User Replaceable | Yes |
Warranty | 2 years |
Waterproof | IP-68 |
Customer Service QQ
Customer Hotline:
Technical Supports