Operational Specifications | ||||||
OUTPUT RANGES (USER SELECTABLE) |
AMBIENT OPERATIONAL TEMPERATURE |
FLOW RATE (RECOMMENDED) |
INLET GAS PRESSURE |
OPERATIONAL CONDITIONS |
||
0 – 10 ppm 0 – 50 ppm 0 – 100 ppm 0 – 500 ppm 0 – 1000 ppm |
0 – 5000 ppm 0 – 1.0% 0 – 5.0% 0 – 10.0% 0 – 25.0% |
41°F to 113°F (5°C to 45°C) |
0.1 to 2.0 SCFH (0.05 – 1.0 Lpm) |
MODEL 201RS 0.5 psig to 15 psig (0.03 bar to 1.0 bar) MODEL 201RSP -6.5 psig (vacuum) to 15 psig (-0.5 bar to 1.0 bar) |
Pollution Degree 2 Installation Category II |
Analyzer Performance Specifications |
90% UPSCALE RESPONSE TIMES FOR SPECIFIC RANGES |
SENSITIVITY | REPEATABILITY | Drift | Humidity | ||
0 – 10 ppm | 0 – 100 ppm | 0 – 1000 ppm | ||||
<25 sec | <10 sec | <10 sec | 0.5% of full scale |
±1.0% of full scale at constant temperature |
±1.0% of full scale in 4 weeks at constant temperature |
<85%, non-condensing |
Analyzer Key Features |
ELIMINATOR CELL BLOCK | Contains a Complete Sample System with all components, including the flow control valve, flow meter, sample, span, and seal-off solenoid valves, fully integrated |
ALARMS |
2 fully adjustable concentration alarms with configurable Alarm logic and Relay
Contacts, featuring: Alarm Bypass Latching/Non-latching Alarm Delays Failsafe/Non-failsafe Trigger Above or Below Setpoint Open or Close on Alarm |
ANALOG OUTPUTS | 4 − 20mA isolated output signals and Modbus bi-directional RS485 communication |
COMMAND CENTER |
Software that works across
AMI’s Analyzers and gives users access to advanced functions: Calibration Details Alarm Configurations Data Logger with Adjustable Interval Setting |