Products

current position:home page>Products>Gas Instrument>AMI>

MODEL 2010BX Trace Oxygen Analyzer

Updated:2023-11-07

Views:1579

  • Brand: AMI
  • Model: 2010BX
  • Description: The MODEL 2010BX is AMI’s most advanced Trace Oxygen Analyzer. This next generation unit combines patented technologies and superior design to deliver unsurpassed performance and reliability while kee
  • Feedback
  • WeChat

    QQ

    Online Service

Product Details

MODEL 2010BX Trace Oxygen Analyzer Description:

Operator Manual Quick Start Guide Data SheetAMI Overview

Features & Benefits

ELIMINATOR CELL BLOCK FOR A LEAK-FREE DESIGN

The patented ELIMINATOR CELL BLOCK provides a unique, compact sample system approach and a flow-efficient sensor pocket machined into a solid metallic block. Precision cross drilled intersecting gas passages eliminate the need for long lengths of tubing and leak-prone compression fittings. Additionally, a uniquely engineered 3-way selector valve (Sample/Span/Off), metering valve, pressure sensor and flow meter are all integrated into the ELIMINATOR CELL BLOCK. It also provides the user with direct front panel access for installing and replacing sensors, and the ability to air calibrate the analyzer, all without the need for disassembly or tools.

PROPRIETARY SENSOR TECHNOLOGY

AMI’s patented oxygen sensors deliver unmatched performance for ultra-fast response times, accuracy, and reliability. Every sensor undergoes an intense battery of tests to ensure they meet our strict quality control standards. Our trace and percent oxygen sensors operate safely, accurately, and reliably in inert or flammable gas streams, even when containing high levels of H2S - up to 500 ppm, for their entire warranted life. This eliminates the need for a maintenance intensive H2S scrubber that competitive oxygen analyzers require.

ALL THE ADVANCED FEATURES FOR ONE LOW PRICE

The MODEL 2010BX Trace Oxygen Analyzer comes loaded with a full suite of advanced features without the typical up charges. The standard features include: 10-user selectable analog output ranges, large backlit LCD, dual independent alarm relay contacts, comprehensive alarm logic (fail safe or non-fail safe, latching or non-latching and independent alarm delays), alarm bypass, and 3 levels of security settings. Automatic data logging provides detailed time-stamped records of O2 measurement, gas pressure, temperature, supply voltage and power brown-out history. All these features can be accessed with AMI’s powerful COMMAND CENTER software, which works universally across all AMI analyzers.

Technical Specifications

Physical & Eletrical Specifications
WEIGHT MOUNTING GAS CONNECTIONS WETTED PARTS
16.0 lbs (7.26 kg) Wall Mount or 2.0” Pipe ¼” 316 S.S. compression fittings 316 S.S. fittings, electro-less nickel-plated cell block, gold-plated contacts, acrylic flow meter & O-rings (Viton, kalrez, and Buna-N)
POWER
DC AC
10 – 24 VDC, 150 mA max (non-heated)
10 – 24 VDC, 2.2 Amps max (heated)
100 – 240 VAC, 150 mA max (non-heated)
100 – 240 VAC, 550 mA max (heated)

Note: 2-year Warranty for Parts & Labor for any defects in materials or workmanship. 6-month Warranty for oxygen sensors

Operational Specifications
OUTPUT RANGES
(USER SELECTABLE)
AMBIENT OPERATIONAL
TEMPERATURE
NON-HEATED Heated HEATED WITH
EXTREME WEATHER
ENCLOSURE
0 – 10 ppm
0 – 50 ppm
0 – 100 ppm
0 – 500 ppm
0 – 1000 ppm
0 – 5000 ppm
0 – 1.0%
0 – 5.0%
0 – 10.0%
0 – 25.0%
25°F to 115°F
(–3.9°C to 46°C)
–20°F to 115°F
(–29°C to 46°C)
–40°F to 115°F
(–40°C to 46°C)
FLOW RATE
(RECOMMENDED)
INLET GAS
PRESSURE
Operational Altitude
DC AC
0.1 to 2.0 SCFH
(0.05 – 1.0 Lpm)
0.5 − 150 psig
(0.03 − 10.3 bar)
≤ 5,500 m
(18,000 ft)
≤ 2,500 m
(8,200 ft)
Sensor & Analyzer Performance Specifications
T-2 Sensor T-4 Sensor 90% UPSCALE RESPONSE TIMES
FOR SPECIFIC RANGES
REPEATABILITY ACCURACY
Max Range
0 –10,000 ppm

Special Conditions
<10 ppm H2S, Up to 50% CO2
Max Range
0 –10,000 ppm

Special Conditions
<500 ppm H2S, Up to 50% CO2
<25 sec for
0.0 – 10 ppm
<10 sec for
10 ppm – 25.0%
±1% of range or
±0.2 ppm of oxygen,
whichever is greater
±1% of full scale range at constant conditions
APPROVALS
Class I, Division 1, Groups B, C & D, T4
Class I Zone 0/1, AEx ia/db IIB+H2 T4 Ga/Gb
IECEX: Ex ia IIB+H2 T4 Ga/Ex db IIB+H2 T4 Gb
UKCA/PESO: Ex ia/db IIB+H2 T4 Ga/Gb
-32°C ≤ Tamb ≤ +46°C
ATEX: II 1/2 G Ex ia/db IIB+H2T4 Ga/Gb
Ingressed Protection: IP65
Analyzer Key Features
ELIMINATOR CELL BLOCK Contains a Complete Sample System with all components, including the flow meter, metering valve, pressure sensor and 3-way Selection (Sample/Span/Off) Valve, fully integrated
ALARMS < fully adjustable concentration alarms with configurable Alarm logic and Relay Contacts, featuring:
Alarm Bypass Latching/Non-latching
Alarm Delays Failsafe/Non-failsafe
Trigger Above or Below Setpoint Open or Close on Alarm
INTERNAL PRESSURE SENSOR Measures gas inlet pressure from 0 – 150 psig
ANALOG OUTPUTS 1 − 5 VDC and 4 − 20mA isolated output signals and Modbus bi-directional RS485 communication
COMMAND CENTER Software that works across AMI’s Analyzers and gives users access to advanced functions:
Error Status Display Alarm Configurations
Data Logger (records O2 measurements, temperature, psig, and power) Calibration Details
ACCESSORIES
Demister
Analyzer Guardian
Extreme Weather Enclosure
Oxygen Sensors

WeChat

Customer Service QQ

Customer Hotline:

010-52867770
0l0-52867771
l3811111452
l7896OO5796

Technical Supports

0l0-52867774
l7896OO5796
info@dorgean.com