Updated:2023-11-07
Views:1579
Online Service
Operational Specifications | ||||
OUTPUT RANGES (USER SELECTABLE) |
AMBIENT OPERATIONAL TEMPERATURE |
|||
NON-HEATED | Heated |
HEATED WITH EXTREME WEATHER ENCLOSURE |
||
0 – 10 ppm 0 – 50 ppm 0 – 100 ppm 0 – 500 ppm 0 – 1000 ppm0 – 5000 ppm 0 – 1.0% 0 – 5.0% 0 – 10.0% 0 – 25.0% |
25°F to 115°F (–3.9°C to 46°C) |
–20°F to 115°F (–29°C to 46°C) |
–40°F to 115°F (–40°C to 46°C) |
FLOW RATE (RECOMMENDED) |
INLET GAS PRESSURE |
Operational Altitude | |
DC | AC | ||
0.1 to 2.0 SCFH (0.05 – 1.0 Lpm) |
0.5 − 150 psig (0.03 − 10.3 bar) |
≤ 5,500 m (18,000 ft) |
≤ 2,500 m (8,200 ft) |
Sensor & Analyzer Performance Specifications | |||||
T-2 Sensor | T-4 Sensor |
90% UPSCALE RESPONSE TIMES FOR SPECIFIC RANGES |
REPEATABILITY | ACCURACY | |
Max Range 0 –10,000 ppm Special Conditions <10 ppm H2S, Up to 50% CO2 |
Max Range 0 –10,000 ppm Special Conditions <500 ppm H2S, Up to 50% CO2 |
<25 sec for 0.0 – 10 ppm |
<10 sec for 10 ppm – 25.0% |
±1% of range or ±0.2 ppm of oxygen, whichever is greater |
±1% of full scale range at constant conditions |
APPROVALS | |||||
Class I, Division 1, Groups B, C & D, T4 Class I Zone 0/1, AEx ia/db IIB+H2 T4 Ga/Gb |
IECEX: Ex ia IIB+H2 T4 Ga/Ex db IIB+H2 T4
Gb UKCA/PESO: Ex ia/db IIB+H2 T4 Ga/Gb -32°C ≤ Tamb ≤ +46°C |
ATEX: II 1/2 G Ex ia/db IIB+H2T4 Ga/Gb Ingressed Protection: IP65 |
Analyzer Key Features | |||||
ELIMINATOR CELL BLOCK | Contains a Complete Sample System with all components, including the flow meter, metering valve, pressure sensor and 3-way Selection (Sample/Span/Off) Valve, fully integrated | ||||
ALARMS |
< fully adjustable concentration alarms with configurable Alarm logic and Relay Contacts, featuring: Alarm Bypass Latching/Non-latching Alarm Delays Failsafe/Non-failsafe Trigger Above or Below Setpoint Open or Close on Alarm |
||||
INTERNAL PRESSURE SENSOR | Measures gas inlet pressure from 0 – 150 psig | ||||
ANALOG OUTPUTS | 1 − 5 VDC and 4 − 20mA isolated output signals and Modbus bi-directional RS485 communication | ||||
COMMAND CENTER |
Software that works across
AMI’s Analyzers and gives users access to advanced functions: Error Status Display Alarm Configurations Data Logger (records O2 measurements, temperature, psig, and power) Calibration Details |
||||
ACCESSORIES |
Demister Analyzer Guardian Extreme Weather Enclosure Oxygen Sensors |
Customer Service QQ
Customer Hotline:
Technical Supports