Products

current position:home page>Products>optical instrument>Ophir>

3A-P 7Z02622 Laser Measurement Sensor

Updated:2022-12-19

Views:1009

  • Brand: OPHIR
  • Model: 3A-P
  • Product description: 3A-P is a very sensitive thermal power/energy laser measurement sensor, which is suitable for short pulse laser. It has a 12 mm aperture. It can measure 15 µ W to 3W and 20 µ J to 2J. It has a P-type
  • Feedback
  • WeChat

    QQ

    Online Service

product details

Overview of 3A-P laser measurement sensor:

3A-P is a very sensitive thermal power/energy laser measurement sensor, which is suitable for short pulse laser. It has a 12 mm aperture. It can measure 15 µ W to 3W and 20 µ J to 2J. It has a P-type volumetric absorber covering a spectral range from 0.15 to 8 µ m. The sensor has a 1.5m long cable for connecting to the instrument or PC interface.

3A-P Sensor Drawing:

Main parameters:

Absorber Type P type
Aperture Ø12mm
Spectral Range 0.15-8µm
Power Range 15µW-3W
Energy Range 20µJ-2J
Dimensions Ø70 W x 30 D (mm)
Max Pulse Energy 2J
Max Average Power Density 0.05kW/cm²
Response Time 2.5 s
Max Energy Density (for <100ns) 1J/cm²
Max Energy Density (for 2ms) 1J/cm²
Max Average Power 3W

Problem summary:

Can the 3A-P sensor operate at 10.6 µm?

The 3A-P sensor actually absorbs about 85% at 10.6 µm, so it can be used to measure weak CO2 lasers.

Can the BeamTrack sensor be used to measure the position and size of the pulsed laser?

Because position and size are measured together with power, if the laser pulses at a rate at which the average power can be measured, then position and size can also be measured.

Must a specific sensor be used only with the instrument used when ordering?

Ophir instruments and sensors are calibrated independently. Each instrument has the same sensitivity as the other instrument, and the error is about 2%. Each sensor is calibrated independently of a specific instrument, and its calibration information is contained in the DB15 plug. When the sensor is connected to the instrument, the instrument will read and interpret this information. Since the accuracy of the sensor is usually+/- 3%, the additional 0.2% error that may result from inserting different instruments can be ignored, so it does not matter which calibration instrument is used for a specific calibrated sensor.

Does the injury threshold depend on the power level?

The damage threshold of the thermal sensor really depends on the power level, not just the power density, because the sensor disk itself will become hotter at high power. For example, the damage threshold of Ophir broadband coating may be 50KW/cm2at 10W, but only 10KW/cm2at 300W. Ophir's damage threshold specification is always given for the maximum power used by a particular sensor.

WeChat

Customer Service QQ

Customer Hotline:

010-52867770
0l0-52867771
l3811111452
l7896OO5796

Technical Supports

0l0-52867774
l7896OO5796
info@dorgean.com