Cera-Diver Groundwater Datalogger Description:

Product Manual Tech Sheet Brochures

Monitoring groundwater under potentially corrosive conditions, such as brackish water and seawater, requires a robust and durable datalogger.

The ceramic-shelled Cera-Diver is designed specifically for such environments. This highly reliable and compact Diver measures groundwater levels with a typical accuracy of ±0.05% full scale.

In addition to its rugged casing, the Cera-Diver’s memory capacity can store up to 48,000 measurements (date/time, pressure and temperature) – almost one measurement every ten minutes for an entire year.

The Cera-Diver has the following sample methods: fixed length intervals, event dependent, averaging and pumping test.

The Cera-Diver is available in the following pressure ranges: 10 m, 20 m, 50 m and 100 m.

Features

Innovative ceramic housing technology

Compact size: Ø 22 mm – length 90 mm

Various measurement methods

48,000 records of time stamp, pressure and temperature

30-Point Pressure Factory Calibration

Benefits

Corrosion proof

Easy installation in almost every well

Event based, averaging and pumping tests sample methods to conserve memory

Cost effective frequent, long-term measuring

Reliable and accurate data

Applications

Saltwater intrusion projects

Leachate monitoring at landfill sites

Monitoring groundwater and surface water

Monitoring stored brines and reclaimed waste water

Estuary and wetlands monitoring

Groundwater monitoring network automation

Aquifer storage and recovery projects

Groundwater monitoring projects

Mining:

Tailings

Environmental

Slope stability

Cera-Diver Groundwater Datalogger Specification:

General
Length 90 mm
Diameter 22 mm
Memory 48,000 measurements
Housing ceramic (ZrO2)
Weight 50 grams
Battery life 10 years (dependent on usage)
Sample interval 0.5 second to 99 hours
Sample methods fixed, event dependent, averaging, and pumping test
Temperature
Value Unit
Range -20 .. +80 °C
Range ±0.1 °C
Resolution 0.01 °C
Pressure
Part No DI701 DI702 DI705 DI710 Range
Range 10 20 50 100 meter
Accuracy ±0.5 ±1.0 ±2.5 ±5.0 cmH2O
Resolution 0.2 0.4 1.0 2.0 cmH2O