Products

current position:home page>Products>Analysis Instrument>PST>

High End Humidity Transmitter - Rotronic HF5

Updated:2023-09-19

Views:698

  • Brand: PST/Rotronic
  • Model: HF5
  • Product description: The HygroFlex5 series is compatible with HC2A probes which offers a high variety in temperature ranges, accuracy and body types. This device generation also boasts a unique calibration and adjustment
  • Feedback
  • WeChat

    QQ

    Online Service

product details

High End Humidity Transmitter - Rotronic HF5 Description:

Product Manual

The HygroFlex5 series is compatible with HC2A probes which offers a high variety in temperature ranges, accuracy and body types. This device generation also boasts a unique calibration and adjustment process as well as many other useful innovations. The HF5 Humidity and Temperature Transmitters are available as wall and duct mount versions.

Exchangeable Probes for Humidity and Temperature Measurement

Features

Compatible with all HC2A probes (interchangeable)

Range of application electronics -40…60 °C / 0…100 %rh

Easy and fast validation through simulator modus

GAMP compatible (HygroSoft Software)

Specifications

HF5
Type 1x Exchangeable HC2A probe
Markets Pharmaceutical Industry
Operating Range -40...60 °C / 0...100 %rh; -10...60 °C with display
Measuring Range -100...200 °C / 0...100 %rh, Depends on Probe
Compatible Probes HC2A-S series, HC2A-IC/IM/IE, HC2-CP, HC2-HK
Probe Cable Extension 5 m passive, up to 100 m with an active cable
Accuracy at 23°C ±0.8 %rh / ± 0.1 K (typical value)
Long-term Stability <1 % RH (typical value), Depends on Probe
Sensors Depends on Probe
Display Version available with and without display
Power Supply HF52 2-wire: 10...28 VDC
HF53 3/4-wire: 15...40 VDC / 12...28 VAC
HF54 electrical isolation 9...36 VDC
HF56 electrical isolation 100...240 VAC"
Output Parameters Humidity and temperature
Calculated Parameters All psychrometric parameters available
Edition 2 analogue outputs 2-wire or 3/4-wire, freely scalable digital outputs optional (LAN / RS-485)
Output Signals 2 wire: 4...20 mA, 3/4-wire: 0(4)...20mA, 0...1/5/10 VDC
Housing Type Wall and Duct Mount
Housing Material ABS
Dimensions wall 160x206 mm with HC2A-S

APPLICATIONS

HVAC applications

Food and pharmaceutical industries

Printing and paper industries

Meteorology

Agriculture

Archaeology

Humidity and Temperature Measurement in Cheese Production

Controlled Atmosphere Monitoring

Coating Process Optimization

WeChat

Customer Service QQ

Customer Hotline:

010-52867770
0l0-52867771
l3811111452
l7896OO5796

Technical Supports

0l0-52867774
l7896OO5796
info@dorgean.com