Updated:2023-09-19
Views:1116
Online Service
The LB 6500 - H10 is a Geiger-Mueller Dose Rate Probe used in various radiation protection applications; dose rate probe for environmental gamma-monitoring or for local dose rate measurement.
The measured quantities are ambient dose equivalent H*(10) and ambient dose rate equivalent Ḣ*(10).
The dose rate probe exists in 2 versions:
1. LB 6500-4-H10 Low dose rate ≤ 10 mSv/h
2. LB 6500-3-H10 High dose rate ≤ 1 Sv/h
LB 6500-4-H10 Specifications
Dose rate range |
100 nSv/h – 10 mSv/h (type approved: 500 nSv/h – 3 mSv/h) |
Energy range | PTB type tested: 65 keV – 1.3 MeV ± 40% with regard to Cs-137 and 0° |
angular response ± 45° |
|
Life cycle | 5 x 1E10 counts |
Intrinsic background | approx. 0.08 cps |
Calibration factor | 0.617 µSv/h per cps |
Operating voltage | 450 Volt |
Operating Conditions | |
Ambient temp. | -20 to +40°C |
Rel. humidity, condensing | 30 to 90 % |
Storage temp. | max. 60°C |
Protection degree | IP65 |
Dimensions / Weight | |
Outer diameter | 45 mm |
Length | 315 mm |
Weight | approx. 1000 g |
LB 6500-3-H10 Specifications
Dose rate range |
1 µSv/h – 1 Sv/h (type approved: 10 µSv/h – 1 Sv/h) |
Energy range | PTB type tested: 50 keV – 1.3 MeV ± 40% with regard to Cs-137 and 0° |
angular response ± 45° |
|
Life cycle | 5 x 1E10 counts |
Intrinsic background | approx. 0.014 cps |
Calibration factor | 7.05 µSv/h per cps |
Operating voltage | 575 Volt |
Operating Conditions | |
Ambient temp. | -20 to +40°C |
Rel. humidity, condensing | 30 to 90 % |
Storage temp. | max. 60°C |
Protection degree | IP65 |
Dimensions / Weight | |
Outer diameter | 45 mm |
Length | 315 mm |
Weight | approx. 1000 g |
APPLICATIONS
Customer Service QQ
Customer Hotline:
Technical Supports