Updated:2023-11-13
Views:1458
Online Service
IDEAL FOR NON-FLAMMABLE GASES
The MODEL 111 Portable Percent Oxygen Analyzer can handle non-flammable samples and is available in 3 configurations: base model, sample pump model and sample pump model for hazardous areas. The Analyzer is used in many applications, including those used for nitrogen purity, inert gas blanketing and oxygen purity.
ZIRCONIUM OXIDE SENSOR TECHNOLOGY
AMI’s zirconium oxide sensors are an extremely stable, have a very long service life, and are not affected by barometric or temperature changes. They are an excellent option for measuring percent levels of oxygen in inert atmospheres, such as the amount of oxygen in the air of an enclosed space. Additionally, these sensors last for many years and require minimal calibration.
FULL OF STANDARD FEATURES
The MODEL 111 Portable Percent Oxygen Analyzer can connect with the COMMAND CENTER software to access advanced features, including automatic data logging that provides detailed time-stamped records of O2 measurement, temperature, supply voltage and power brown-out history.
TRULY PORTABLE AND DESIGNED FOR THE FIELD
Starting at 5 lbs (7.3 kg), the MODEL 111 is the smallest, lightest, fastest and most accurate portable Percent Oxygen Analyzer in the industry.
Operational Specifications | ||||||
OUTPUT RANGES (USER SELECTABLE) |
AMBIENT OPERATIONAL TEMPERATURE |
FLOW RATE (RECOMMENDED) |
INLET GAS PRESSURE |
|||
Standard | Optional | Model 111 | Model 111P | Model 111B | ||
0 – 1.0% 0 – 5.0% 0 – 10.0% 0 – 25.0% |
0 – 50.0% 0 – 95.0% |
0°F to 130°F (-18°C to 54°C) |
0.1 to 2.0 SCFH (0.05 – 1.0 Lpm) |
0.5 − 100 psig (0.03 − 6.9 bar) |
-14 in Hg − 10 psig (0.03 − 0.7 bar) |
-14 in Hg − 10 psig (0.03 − 0.7 bar) |
Analyzer Performance Specifications | |||||
90% UPSCALE RESPONSE TIMES FOR SPECIFIC RANGES |
REPEATABILITY | APPROVALS | |||
0 - 25% | 0 - 95% | Model 111 | Model 111P | Model 111B | |
< 12 sec | < 12 sec |
±0.1% of range or ±0.1% of oxygen, whichever is greater |
Meets UL Requirements (EN 61326-1). CE Mark |
Meets UL Requirements (EN 61326-1). CE Mark |
Designed to meet Class 1, Div 2, Groups C,D |
Analyzer Key Features | ||
ZIRCONIUM OXIDE SENSOR | 10-year life expectancy (note: sensor is not field replaceable) | |
COMMAND CENTER |
Software that works across AMI’s
Analyzers and gives users access to advanced functions: Error Status Display Calibration Details
Time-stamped data logger with adjustable
interval recording for:
• O2 Measurement • Temperature • Battery Voltage |
|
ACCESSORIES |
Carrying Case Flexible High Pressure Sample Tubing Auxiliary Power Adapter |
Customer Service QQ
Customer Hotline:
Technical Supports