Products

current position:home page>Products>optical instrument>Ophir>

3A 7Z02621 Laser Power/Energy Sensor

Updated:2022-12-19

Views:1466

  • Brand: OPHIR
  • Model: 3A
  • Description: 3A is a very sensitive laser power/energy sensor based on thermopile, with very low noise and drift. It has a 9.5mm aperture. It can measure power from 10 µ W to 3W and from 20 µ J to 2J. It has a fla
  • Feedback
  • WeChat

    QQ

    Online Service

Product Details

Overview of 3A laser power/energy sensor:

3A is a very sensitive thermopile based laser power/energy sensor with very low noise and drift. It has a 9.5mm aperture. It can measure from 10µW to 3W and from 20µJ to 2J. It is spectrally flat and covers the spectral range from 0.19 to 20 µm. The sensor comes with a 1.5 meter cable for connecting to a meter or PC interface.

3A Sensor Drawing:

Main parameters:

Absorber Type Low power broadband
Aperture Ø9.5mm
Spectral Range 0.19-20µm
Power Range 10µW-3W
Energy Range 20µJ-2J
Dimensions Ø70 W x 30 D (mm)
Max Pulse Energy 2J
Max Average Power Density 1kW/cm²
Response Time 1.8 s
Max Energy Density (for <100ns) 0.3J/cm²
Max Energy Density (for 2ms) 2J/cm²
Max Average Power 3W

Problem summary:

What does it mean that the spectral range exceeds the calibrated wavelength range?

The spectral range described at the beginning of the specification represents the range of wavelengths that the sensor can effectively use, even if no precise calibration is specified for that range. This means that the accuracy is specified and guaranteed within the calibrated wavelength range. In a wider range of useful wavelengths, the sensor can be used but its accuracy cannot be guaranteed. Generally speaking, within this wider range, the accuracy will be within ± 15%.

Must a specific sensor be used only with the instrument used when ordering?

Ophir instruments and sensors are calibrated independently. Each instrument has the same sensitivity as the other instrument, and the error is about 2%. Each sensor is calibrated independently of a specific instrument, and its calibration information is contained in the DB15 plug. When the sensor is connected to the instrument, the instrument will read and interpret this information. Since the accuracy of the sensor is usually+/- 3%, the additional 0.2% error that may result from inserting different instruments can be ignored, so it does not matter which calibration instrument is used for a specific calibrated sensor.

Does the injury threshold depend on the power level?

The damage threshold of the thermal sensor really depends on the power level, not just the power density, because the sensor disk itself will become hotter at high power. For example, the damage threshold of Ophir broadband coating may be 50KW/cm2at 10W, but only 10KW/cm2at 300W. Ophir's damage threshold specification is always given for the maximum power used by a particular sensor.

WeChat

Customer Service QQ

Customer Hotline:

010-52867770
0l0-52867771
l3811111452
l7896OO5796

Technical Supports

0l0-52867774
l7896OO5796
info@dorgean.com