Products

current position:home page>Products>Measuring Instrument>Campbell>

CS650 30 cm Soil Moisture and Temperature Sensor

Updated:2023-11-29

Views:734

  • Brand: Campbell
  • Model: CS650
  • Product description: The CS650 is a multiparameter smart sensor that uses innovative techniques to monitor soil volumetric water content, bulk electrical conductivity, and temperature. It outputs an SDI-12 signal that man
  • Feedback
  • WeChat

    QQ

    Online Service

product details

CS650 30 cm Soil Moisture and Temperature Sensor Description:

Product Manual Datasheet

The CS650 is a multiparameter smart sensor that uses innovative techniques to monitor soil volumetric water content, bulk electrical conductivity, and temperature. It outputs an SDI-12 signal that many of our data loggers can measure.

Detailed Description

The CS650 consists of two 30-cm-long stainless steel rods connected to a printed circuit board. The circuit board is encapsulated in epoxy and a shielded cable is attached to the circuit board for data logger connection.

The CS650 measures propagation time, signal attenuation, and temperature. Dielectric permittivity, volumetric water content, and bulk electrical conductivity are then derived from these raw values.

Measured signal attenuation is used to correct for the loss effect on reflection detection and thus propagation time measurement. This loss-effect correction allows accurate water content measurements in soils with bulk EC ≤3 dS m-1 without performing a soil specific calibration.

Soil bulk electrical conductivity is also calculated from the attenuation measurement. A thermistor in thermal contact with a probe rod near the epoxy surface measures temperature. Horizontal installation of the sensor provides accurate soil temperature measurement at the same depth as the water content. Temperature measurement in other orientations will be that of the region near the rod entrance into the epoxy body.

Benefits and Features

More accurate water content measurements in soils with bulk EC up to 3 dS m-1 without performing a soil-specific calibration

Larger sample volume reduces error

Measurement corrected for effects of soil texture and electrical conductivity

Estimates soil-water content for a wide range of mineral soils

Versatile sensor—measures dielectric permittivity, bulk electrical conductivity (EC), and soil temperature

CS650 30 cm Soil Moisture and Temperature Sensor Specifications:

Measurements Made Soil electrical conductivity (EC), relative dielectric permittivity, volumetric water content (VWC), soil temperature
Required Equipment Measurement system
Soil Suitability Long rods with large sensing volume (> 6 L) are suitable for soils with low to moderate electrical conductivity.
Rods Not replaceable
Sensors Not interchangeable
Sensing Volume 7800 cm3 (~7.5 cm radius around each probe rod and 4.5 cm beyond the end of the rods)
Electromagnetic CE compliant
Meets EN61326 requirements for protection against electrostatic discharge and surge.
Operating Temperature Range -50° to +70°C
Sensor Output SDI-12; serial RS-232
Warm-up Time 3 s
Measurement Time 3 ms to measure; 600 ms to complete SDI-12 command
Power Supply Requirements 6 to 18 Vdc (Must be able to supply 45 mA @ 12 Vdc.)
Maximum Cable Length 610 m (2000 ft) combined length for up to 25 sensors connected to the same data logger control port
Rod Spacing 32 mm (1.3 in.)
Ingress Protection Rating IP68
Rod Diameter 3.2 mm (0.13 in.)
Rod Length 300 mm (11.8 in.)
Probe Head Dimensions 85 x 63 x 18 mm (3.3 x 2.5 x 0.7 in.)
Cable Weight 35 g per m (0.38 oz per ft)
Probe Weight 280 g (9.9 oz) without cable
Current Drain
Active (3 ms)
  • 45 mA typical (@ 12 Vdc)
  • 80 mA (@ 6 Vdc)
  • 35 mA (@ 18 Vdc)
Quiescent 135 µA typical (@ 12 Vdc)
Electrical Conductivity
Range for Solution EC 0 to 3 dS/m
Range for Bulk EC 0 to 3 dS/m
Accuracy ±(5% of reading + 0.05 dS/m)
Precision 0.5% of BEC
Relative Dielectric Permittivity
Range 1 to 81
Accuracy
  • ±(2% of reading + 0.6) from 1 to 40 for solution EC ≤ 3 dS/m
  • ±1.4 (from 40 to 81 for solution EC ≤1 dS/m)
Precision < 0.02
Volumetric Water Content
Range 0 to 100% (with M4 command)
Water Content Accuracy
  • ±1% (with soil-specific calibration)
  • ±3% (typical with factory VWC model) where solution EC < 3 dS/m
Precision < 0.05%
Soil Temperature
Range -50° to +70°C
Resolution 0.001°C
Accuracy
  • ±0.1°C (for typical soil temperatures [0 to 40°C] when probe body is buried in soil)
  • ±0.5°C (for full temperature range)
Precision ±0.02°C

WeChat

Customer Service QQ

Customer Hotline:

010-52867770
0l0-52867771
l3811111452
l7896OO5796

Technical Supports

0l0-52867774
l7896OO5796
info@dorgean.com