Products

current position:home page>Products>Measuring Instrument>Apogee>

MQ-501: Handheld Full-spectrum Quantum Meter

Updated:2022-12-27

Views:980

  • Brand: Apogee
  • Model: MQ-501
  • Description: MQ-501 is designed as an easy to use hand-held instrument for spot check and measurement. The instrument includes AM-001 instrument mounting bracket for mounting the sensor on the horizontal surface o
  • Feedback
  • WeChat

    QQ

    Online Service

Product Details

Overview of MQ-501 hand-held full spectrum quantum meter:

User 's Manual

MQ-501 is designed as an easy to use hand-held instrument for spot check and measurement. The instrument includes AM-001 instrument mounting bracket for mounting the sensor on the horizontal surface of the instrument, and approximately 0.17 m of cable to accommodate the length of the bracket. The full spectrum sensor adopts blue enhanced silicon photodiode and customized filter, with a solid self-cleaning sensor housing design, anodized aluminum body and acrylic diffuser. Typical applications include PPFD measurement of plant canopy in outdoor environment, greenhouse and growth chamber, and PPFD measurement of reflection or under canopy (transmission) in the same environment.



The instrument has sampling and logging modes and will record in mol m ˉ² d ˉ¹ Is the comprehensive daily total amount in units. The sampling mode will record up to 99 manual measurements. The logging mode will turn the instrument on/off and take measurements every 30 seconds. Every 30 minutes, the instrument will average 60 30 second measurements and record the average value into memory. The instrument can store up to 99 average values. Once it is full, it will begin to overwrite the old measured values with new ones. The combined daily total will be recorded from 48 average measurements (over a 24-hour period). The sample and log measurements can be viewed on the LCD screen or by downloading data to the computer, but the combined daily totals can only be viewed by downloading data to the computer. AC-100 communication cable (standard USB cable does not work) and Apogee AMS software are required to download data to the computer.

MQ-501 drawing:


MQ-501 specification:

Calibration Uncertainty ± 5 %
Measurement Range 0 to 4000 µmol mˉ² sˉ¹
Measurement Repeatability Less than 0.5 %
Long-term Drift (Non-stability) Less than 2 % per year
Non-linearity Less than 1 % (up to 4000 µmol mˉ² sˉ¹)
Response Time Less than 1 ms
Field of View 180˚
Spectral Range 389 to 692 nm ± 5 nm (wavelengths where response is greater than 50 % of maximum)
Directional (Cosine) Response ± 5 % at 75 ˚ zenith angle
Azimuth Error Less than 0.5 %
Tilt Error Less than 0.5 %
Temperature Response -0.11 ± 0.03 % Cˉ¹
Uncertainty in Daily Total Less than 5 %
Detector Blue-enhance silicon photodiode
Housing Anodized aluminum body with acrylic diffuser
IP Rating IP68
Operating Environment 0 to 50 ℃, less than 90 % non-condensing relative humidity up to 30 C, less than 70 % non-condensing relative humidity from 30 to 50 ℃, separate sensors can be submerged in water up to depth of 30 m
Meter Dimensions 126 mm length, 70 mm width, 24 mm height
Sensor Dimensions 24 mm diameter, 37 mm height
Mass 180 g
Cable Approximately 0.17 m of shielded, twisted-pair wire with TPR jacket (high water resistance, high UV stability, flexibility in cold conditions)
Warranty 4 years against defects in materials and workmanship
Manufactured Made in the USA

WeChat

Customer Service QQ

Customer Hotline:

010-52867770
0l0-52867771
l3811111452
l7896OO5796

Technical Supports

0l0-52867774
l7896OO5796
info@dorgean.com