Oxygen Analyzer - Michell XTP601

Updated:2023-10-20

Views:1063

  • Brand: PST/Michell
  • Model: XTP601
  • Description: The XTP601 Oxygen Analyzer for safe or hazardous areas is a robust, linear and stable device that is used for measurements in gases such as biogas, methane, hydrogen, nitrogen or carbon dioxide. The a
  • Feedback
  • WeChat

    QQ

    Online Service

Product Details

Oxygen Analyzer - Michell XTP601 Description:

User’s Manual Datasheet Order Selection Table

Low-maintenance O2 Measurement

Features

General purpose or ATEX, UKCA, IECEx, cQPSus, TR CU & Ex hazardous area certified models

SIL2 Capable

Ranges available: 0...0.5 % up to 0...50 % & 20, 80 or 90...100 %

Touch-screen display allows calibration or adjustment without the need for a hot works permit

Intrinsic error (accuracy) is better than 1 % full scale

Zero range stability better than 0.25 % of range per month

Outputs include RS485 and two 4...20 mA (following NAMUR NE 44 standard)

Low cost of ownership due to minimal maintenance and no consumable parts

The XTP601 Oxygen Analyzer for safe or hazardous areas is a robust, linear and stable device that is used for measurements in gases such as biogas, methane, hydrogen, nitrogen or carbon dioxide. The analyzer is SIL2 Capable.

A robust, linear and stable oxygen analyzer for measurements in gases such as hydrogen, nitrogen, methane, biogas, or carbon dioxide. The analyzer is housed in an IP66 rugged casing, making it suitable for a wide range of applications. The XTP601 can be supplied with flame arrestors, as either explosion-proof and certified for use in hazardous areas or for use with flammable gases in a safe area.

Specifications

Performance
Measurement Technology Thermo-paramagnetic oxygen sensor
Gas Process and non-condensing sample with particles <5µm
Measurement Range Selectable from 0 to 0.5 % up to 0...50 % and 20, 80 or 90...100 %
Display Resolution 0.01 % for zero based ranges
0.1 % for supressed zero ranges
Display Type Backlit LCD
Accuracy Excluding Suppressed Zero Ranges ±0.02 % O2 or <±1 % of range, whichever is greater
Accuracy For Suppressed Zero Ranges (80/90...100 %) ±0.2 % O2 or <±1 % of range, whichever is greater
Response Time (T90) with High Speed Response (HSR) Enabled < 15 seconds
Repeatability ±0.2 % of range or 0.01 % O2, whichever is greater
Linearity ±0.5 % of range or 0.05 % O2, whichever is greater
Zero Stability ±0.25 % of range per month
Range Stability ±0.25 % of range per month
Sample Flow Rate 100...600 ml/min
Sample Flow Effect GP1 and GP3: 100...500 ml/min (0.2...1.06 scfh)
EX1, EX3 and GP2: 270...330 ml/min (0.2...1.06 scfh)
Sample Pressure 0.75...2 BarA
Sample Temperature A constant temperature of +5...+45/+55 °C
(+41...+113/+131 °F)
Sample Cell Temperature +50/+60 °C (+122/+140 °F)
Background Gas Analyzer is calibrated in the background gas of the process. Standard gases are: N2, CO2, CH4, H2 & Biogas. Others available on request
Electrical Specifications
Analog Inputs 2 off 4...20 mA inputs
One for an external sensor that can be displayed on the screen
One to act as an active compensation for the process conditions
Analog Outputs 2 off 4...20 mA outputs (powered with 24V excitation voltage)
Output Ranges Primary range is set to the calibrated range of the instrument
The second is user configurable between 0 and 100 %
Alarms 2 off single pole changeover (SPCO) relays for O2 concentration (250 V, 5 A max)
Datalogging The user can use the digital communications to log the output from the analyzer
The unit will store 40 alarm points and the min/max O2 concentrations with date and time stamp
Digital Communications Modbus RTU over RS485
Power Supply 24 V DC; 1.6 A max
Cable entries Analyzer is supplied with 3 x M20 cable entries. Cable glands, conduit entries and blanking plugs are available as accessories
Operating Conditions
Ambient Temperature +5...+60 °C (+41 °F...+140 °F)
+5...+50 °C (+41...+122 °F) cQPSus
Atmopheric pressure 750 mbar...1250 mbar
Mechanical Specifications
Warm Up Time <25 minutes
Stabilization Time 5 minutes
Dimensions 234 x 234 x 172 mm (w x d x h)
Weight 9.7 kg
Wetted Materials 316 & 430F stainless steel, borosilicate glass, platinum, 3M 2216, plus O-ring
O-Ring Materials Viton, Silicone or Ekraz
Gas Connection EX1, EX3 & GP2: 1/8” NPT (female)
GP1 and GP3: 1/4” NPT (female)
Ingress Protection IP66, NEMA 4X
Hazardous Area Classification
ATEX/UKCA II 2 G D Ex db IIB+H2 T6 Gb
Ex tb IIIC T85 °C Db IP66
IECEx Ex db IIB+H2 T6 Gb
Ex tb IIIC T85 °C Db IP66
cQPSus Class I, Division 1, Groups B,C,D, T6
Class I, Zone 1
AEx db IIB+H2 T6 Gb / Ex db IIB+H2 T6 Gb
TR CU 1 Ex d IIB+H2 T6 Gb
Temperature Ranges for ATEX/UKCA and IECEx
as per O-ring Type
Silicon: Ta = -40 °C...+60 °C (+50 °C for N. America, +55 °C for TR CU Ex (EAC)).
Viton®: Ta = -15 °C...+60 °C (+50 °C for N. America, +55 °C for TR CU Ex (EAC))
Ekraz: Ta = -10 °C...+60 °C (+50 °C for N. America, +55 °C for TR CU Ex (EAC))