Fast Responding With Metallic Sealed Reference Sensor
Features
Fast response time
Simple and easy operation
Low maintenance and cost of ownership
No need for reference air
Barometric pressure compensation
Multiple Outputs available from 4-20 mA, RS485 Modbus RTU and RS232
Rack, wall-mount, bench-mount, and transportable versions available
Optional built-in Easidew Transmitter with rack-mount version
The XZR400 zirconium oxide gas purity analyzer from Michell Instruments is designed to measure oxygen as an impurity in inert gases such as nitrogen, carbon dioxide, helium and argon.
A range of trace oxygen analyzers utilizing Michell's Metallic Sealed Reference Sensor (MSRS) technology. Designed to measure O2 as an impurity in pure inert gases such as N2, CO2, He, Ar and even Kr. The XZR400 can also be used for cylinder filling and other applications where a precise measurement of oxygen is required. There are three different chassis to choose from and various options including pumps, flow alarms and digital communications to match customer needs. The rack mount unit can be supplied with an Easidew or SF82 moisture transmitter fitted inside the case.
Specifications
Sensor Type | |
Measurement Principle | Zirconium Oxide Sensor with Metallic Sealed Reference and S Type thermocouple |
Performance | |
Gas | Clean, dry, oil free with particles less than 3μm |
Measurement Range | 0.1 ppm up to 25% O2 Extended ranges available up to 0 to 100% O2 |
Lowest detectable limit (LDL) | 0.1ppm(v) O2 |
Accuracy (Intrinsic Error) | Less than 2% of reading |
Response Time | <11 seconds |
Repeatability | ±0.1% of reading |
Stability | 1% per month |
Linearity | Better than ±1% |
Drift | <1% of reading per week |
Sample Flow Rate | 1 to 3 Nl/h with built-in fast loop |
Maximum Sample Pressure | 2 barg(29 psig) |
Maximum Sample Temperature | +100°C (+212°F) |
Atmospheric Pressure Compensation | Built-in as standard |
Optional sensor | |
Moisture sensor | Easidew sensor can be fitted as an option. Please, see a separate datasheet for full specifications. |
Measurement range | -100 °C to + 20 °C |
Sample Flow rate | 60 to 300 Nl/hour |
Outputs | |
Analog Output | 0, 1 or 2 off 0/4-20 mA Linear with Galvanic Isolation output |
Digital Communications Output |
Modbus RTU over RS485 protocol (Standard on XZR400A1) RS232 output (optional) |
Output Load | Over 1000 Ω |
Self-diagnostics | Via HMI |
Output Ranges |
0.0 ppm to 1ppm up to 0 to 25% O2 Extended ranges available up to 0 to 100% O2 |
Alarms |
2 threshold alarms, freely configurable 1 general fault alarm including flow alarm 1 flowalarm (optional) |
Display Resolution |
0.01 ppm between 0.1 ppm and 10 ppm 0.1 ppm between 10 ppm and 10,000 ppm 0.01% between 1 to 10% 0.1% between 10 to 25% |
Power Supply | 90 to 264 V AC, 47/63Hz |
Power Consumption | 50 VA |
Operating Conditions | |
Ambient Temperature Range | 0 to +55°C (+32 to +131°F) |
Sensor Temperature | Optimized @ +634°C |
Operating Humidity | 5 to 90% RH without condensation |